Temperature and humidity outdoor, indoor probe with 0-10V output

Ambient temperature, relative humidity transmitter. Outdoor, indoor use.

Built-in relative humidity, temperature sensors. Measured values are also converted to other humidity interpretation: dew point temperature, absolute humidity, specific humidity, mixing ratio, specific enthalpy.

NEW - AC Power -  now also powered from 24 Vac !

  • code
    T0210
  • Warranty
    3 years

Technical data

Measured value

Temperature
Relative humidity
Counted values (dew point…)
Relative humidity range 0 to 100%
Accuracy of relative humidity measurement ±2.5% relative humidity from 5 to 95% at 23°C
Accuracy of temperature output ±0.4°C
Available temperature units degrees Celsius, Fahrenheit
Accuracy and range of dew point temperature output - for more details see graphs ±1.5°C at ambient temperature T<25°C and RH>30%
range -60 to +80 °C
Accuracy and range of absolute humidity output ±1.5g/m3 at ambient temperature T < 25°C
range 0 to 400 g/m3
Accuracy and range of specific humidity output ±2g/kg at ambient temperature T < 35°C
range 0 to 550 g/kg
Accuracy and range of mixing ratio output ±2g/kg at ambient temperature T < 35°C
range 0 to 995 g/kg
Accuracy and range of specific enthalpy output ±3kJ/kg at ambient temperature T < 25°C
range: 0 to 995 kJ/kg
Temperature operating range -30 to +80°C
Temperature operating range of LCD display readable to operating temperature +70°C, it is recommended to switch OFF the LCD over +70°C
Range of humidity sensor temperature compensation all temperature range
Voltage outputs 0-10V, outputs are not galvanic isolated - have common ground
Configuration of outputs and output range user adjustable from the PC
Filtering ability of sensor cover 0.025mm - filter with stainless steel mesh
Protection of the case with electronics IP65 electronics, IP40 sensors
Power 15-30Vdc, maximum consumption 20mA; 24 Vac
Dimensions 88.5 x 170 x 39.5 mm (W x H x D), stem length 75 mm
Weight approximately 150g
Warranty 3 years