This website is using cookie files for provision of services, personalization of advertising and for analysing of visit rate.
By using this site you agree to our use of cookies. More information

manufacturer of measuring instruments
English Česky USA Polski Sweden Africa Magyar Español China Français

Thermometer hygrometer barometer with Ethernet interface and relays

code: H7530

Ambient temperature, relative humidity, atmospheric pressure with two relay outputs. Humidex reading.


Sensor H7530 is designed for online monitoring of temperature, relative humidity of air without aggressive substances. Three binary inputs to detection of two-states signals are the advantage. Other devices are controlled by two relay outputs.

High precision capacitive polymer sensor ensures excellent long term calibration stability and ultimate accuracy. Measured values are also converted to other humidity interpretation: dew point temperature, absolute humidity, specific humidity, mixing ratio and specific enthalpy.

Included in delivery:

 


Send request

Technical data

Value
Measured value Temperature
Relative humidity
Counted values (dew point…)
Atmospheric pressure
Relative humidity range 0 to 100%
Accuracy of relative humidity measurement ±2.5% relative humidity from 5 to 95% at 23°C
Accuracy of temperature output ±0.4°C
Maximum measuring pressure range 600 to 1100hPa
Accuracy of atmospheric pressure ±1.3hPa at 23°C
Resolution 0.1°C, 0.1%RH, 0.1hPa
Measuring interval 2s
Available temperature units degrees Celsius, degrees Fahrenheit
Computed values humidex, dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy
Available pressure units hPa, kPa, mbar, mmHg, inHg, inH2O, PSI, oz/in<sup>2</sup>
Accuracy and range of dew point temperature output - for more details see graphs in manual ±1.5°C at ambient temperature T<25°C and RH>30%<br />range -60 to +80 °C
Accuracy and range of absolute humidity output ±1.5g/m<sup>3</sup> at ambient temperature T < 25°C<br />range 0 to 400g/m<sup>3</sup>
Accuracy and range of specific humidity output ±2g/kg at ambient temperature T < 35°C<br />range 0 to 550g/kg
Accuracy and range of mixing ratio output ±2g/kg at ambient temperature T < 35°C<br />range 0 to 995g/kg
Accuracy and range of specific enthalpy output ±3kJ/kg at ambient temperature T < 25°C<br />range: 0 to 995kJ/kg
Temperature compensation of the humidity sensor all temperature range
Temperature operating range -30 to +80°C
IP protection IP40
Number of relay outputs 2
Maximum switching voltage, current and power of relay output 50V, 2A, 60VA
Number of binary inputs 3
Signal for binary inputs dry contact, open collector or two-state voltage signal.<br />Inputs are not galvanically isolated.
Minimum pulse length at binary input 500ms
Low voltage level 0 to +0.5V
High voltage level +3.0 to +30V
Audible alarm built-in beeper - switchable
LAN connection RJ-45 connector, 10Base-T or 100Base-TX
Communication protocols WWW, ModbusTCP, SNMPv1, SOAP, XML
Alarm protocols E-mail (SMTP authentication is supported), SNMP Trap, Syslog
Configuration WWW interface, T-Sensor
Temperature operating range of LCD display readable to operating temperature +70°C, it is recommended to switch OFF the LCD over +70°C
Power 9 to 30Vdc, maximum consumption 1W
Power connector co-axial, diameter 5.5 x 2.1mm
Dimensions 136 x 213 x 45mm (W x H x D), stem length 75mm
Weight approximately 360g
Warranty 3 years

Features

Applications:

Aplikace - Sklady

Production facilities and warehouses - Monitoring of storage conditions and production processes in the temperature range -200 ° C to + 600 ° C. Sending data to COMET Database or SCADA third-party systems.

Aplikace - Serverovny a datová centra

Data centers, server rooms - Temperature and humidity monitoring, data center monitoring, RACK stands, including flood status reports, opening / closing of doors (windows), motion, smoke. SNMP monitoring, remote alarms by email and Syslog.

Aplikace - Zdravotnictví a laboratoře

Healtcare and laboratories - -200 °C.

Aplikace - Průmysl

Industry and production - monitoring temperature and humidity in food, pharmaceutical, aerospace etc.

Aplikace - Sklady

Food industry - Monitoring of critical processes according to the HACCP standard with the possibility of prompt notification of unforeseen events that could lead to the degradation of the goods.

 

Features description:

* the features may vary depending on the device type
● accurate measurement of Temperature, Relative humidity, Dew point, Atmospheric pressure, CO2, Two-state events
● communication protocols SNMP, Modbus TCP, SOAP, XML
● webserver to display current and recorded values
● alarm indication via email
● selected models with PoE
● data analysis in COMET Database software
 
Záruka 3 roky Kalibrační listComet Database Teplotní čidlo Teplotní čidlo Teplotní čidlo Tlakové čidlo Rosný bod Humidex Reléové výstupy
 
Dvouřádkový LCD displej Akustická signalizace Ethernetové rozhraní Power over Ethernet - PoE WWW server WWW server Historie uložených hodnot Historické grafy E-mail Export historických dat do CSV
 
Modbus TCP protokol Aktuální hodnoty pomocí XML SNMP protokol SNMP Trap SOAP Protokol Syslog protokol SNTP protokol - časová synchronizace Paměť MIN/MAX hodnot
 

Online demo